Contact an Account ManagerHPLC method for the determination of vitamin E in serum
Summary
Type: HPLC
Application #: 109710
Catalog#: 720014.40 - MN HPLC Column, Nucleosil C18, 100A, 5um, 4x250mm
Detection: A: UV, 284nm
B: fluorescence, excitation 292nm, emission 325nm
Conditions: A)
eluent: acetonitrile - methanol - water (65:34:1, v/v)
flow rate 2ml/min
B)
eluent: methanol
flow rate 1.5ml/min
Authors: Seitz S
Source: J Anal Chem 343 (1992) 77 - 78